Measurement of air refractive index based on surface plasmon resonance and phase detection
Qianghua Chen1, Huifu Luo, Sumei Wang
1Key Laboratory of Advanced Machining Fundamental Science, School of Mechanical Engineering, Beijing Institute of Technology, Teaching Building 1, 5 South Zhongguancun Street, Haidian District, Beijing 100081, China. chenqianghua@tsinghua.org.cn
Abstract:
A method for refractive index of air measurement is presented based on surface plasmon resonance (SPR) and phase detection using a dual-frequency laser interferometer. Theoretical analyses indicate that the phase-difference variation of the measurement signal versus the reference signal is linear with refractive index of air (RIA) fluctuation, and the calculation formula of RIA is derived. The structure design of the self-adaptive SPR sensor greatly reduces the measurement error resulting from the incident angle shift and improves the sensitivity. The experiments show that measurement uncertainty of 10(-6) order has been achieved when phase detection precision is 0.1°. The phenomenon of sudden phase variation during air pumping and air filling, which is caused by temperature fluctuation, is discussed.