Staff doses in interventional radiology: a national survey
Roberto Mariano Sánchez1, Eliseo Vano, Jose M Fernández
1Medical Physics Department, Hospital Clínico San Carlos, Instituto de Investigación Sanitaria del Hospital Clínico San Carlos, Spain. rmsanchez.hcsc@salud.madrid.org
Purpose:
To present the results of occupational radiation doses investigated through a national survey promoted by the National Society of Interventional Radiology in Spain.
Materials And Methods:
The monthly dosimetric records of 28 interventional radiologists from 10 hospitals were analyzed and filtered to remove inconsistent dosimeter readings. The evaluation of the results includes different workloads as well as different radiation protection habits.
Results:
Poor use of personal dosimetry by some interventional radiologists was brought to light. Most professionals do not use an over-apron dosimeter as recommended by the International Commission on Radiological Protection. Ceiling-suspended protective screens are used irregularly in many cases. All interventionalists perform digital subtraction angiographic imaging from a control room in more than 80% of procedures. The maximum monthly doses recorded were 3.8 mSv under the apron, 20.2 mSv over the apron, and 63.1 mSv to the hands.
Conclusions:
For under-apron and hand readings, extrapolated median values were below 30% of annual dose limits, but in the case of over-apron readings, the extrapolated median dose was higher than the newly recommended limit for the eye lens of 20 mSv per year. This study mainly highlights the need to use radiation protection tools and personal dosimeters to protect staff and monitor eye lens doses.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
05:04Using Simulation Models to Train Clinicians in the Use of Point-of-Care Ultrasound
Published on: August 9, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...