Transport with ongoing resuscitation: a comparison between manual and mechanical compression
Holger Gässler1, Marc-Michael Ventzke, Lorenz Lampl
1Department of Anesthesiology & Intensive Care Medicine, Section Emergency Medicine, Armed Forces Medical Centre Ulm, Ulm, Germany. holgergaessler@bundeswehr.org
External mechanical chest compression devices offer a safer alternative during out-of-hospital cardiac arrest patient transport. While some devices ensure constant compressions, only the animax mono met guideline standards for frequency and depth.
Area of Science:
- Emergency Medicine
- Cardiovascular Research
- Biomedical Engineering
Background:
- Transporting out-of-hospital cardiac arrest patients with ongoing resuscitation presents unique challenges.
- External mechanical chest compression devices are being explored as alternatives to manual compressions during transport.
Purpose of the Study:
- To compare the effectiveness of manual chest compressions versus mechanical devices (LUCAS 2, AutoPulse, animax mono) during simulated ambulance transport.
- To evaluate compression frequency, depth, and consistency under transport conditions.
Main Methods:
- A resuscitation manikin (Ambu Man Wireless MegaCode) was used for 10 trials per device.
- Comparisons included manual compressions, LUCAS 2, AutoPulse, and animax mono.
- Testing was conducted in an ambulance vehicle over a 5.0 km track.
Main Results:
- Manual and animax mono achieved higher compression frequencies than LUCAS 2 and AutoPulse.
- Mechanical devices (LUCAS 2, AutoPulse, animax mono) provided constant compressions, unaffected by vehicle motion.
- Only animax mono met recommended compression depth (51.2 mm); manual compression quality degraded during maneuvers.
Conclusions:
- Mechanical chest compression devices may enhance rescuer and patient safety during resuscitation transport.
- Animax mono was the only device to meet guideline specifications for compression frequency and depth.
- Mechanical devices offer reliable, motion-independent compressions, but further in-vivo studies are needed to confirm patient transport effectiveness.