Objectives:
The aim of this study was to find the electromagnetic interference (EMI) thresholds for several commonly used implantable cardioverter-defibrillators (ICD).
Design:
Seventeen ICDs were exposed to magnetic fields with different intensities produced by the Helmholtz coil system. Sinusoidal, pulse, ramp, and square-waveforms with a frequency range of 2 Hz to 1 kHz were used.
Results:
ICD malfunctions occurred in 11 of the 17 ICDs tested. The ICD malfunctions that occurred were false detections of ventricular tachycardia (6/17 ICDs) and ventricular fibrillation (3/17 ICDs), false detection of atrial tachycardia (4/6 dual chamber ICDs) and tachycardia sensing occurring during atrial or ventricular refractory periods (1/17 ICD). In most cases, no interference occurred at magnetic field levels below the occupational safety limits of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). Nevertheless, some frequencies using sine, ramp or square waveforms did interfere with certain ICDs at levels below these limits. No EMI occurred with any of the ICDs below the ICNIRP limits for public exposure.
Conclusion:
Evaluation of EMI should be part of the risk assessment of an employee returning to work after an ICD implantation. The risk assessment should consider magnetic field intensities, frequencies and waveforms.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
11:28Concurrent EEG and Functional MRI Recording and Integration Analysis for Dynamic Cortical Activity Imaging
Published on: June 30, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
Mechanism of Cardiac Arrhythmias
Holter Monitor: 24-Hour Monitoring
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias