An international proficiency study with the tumor marker CA 125
M Zwirner1, C H Bieglmayer, R Klapdor
1Dept. of Obstetrics and Gynecology, University of Tübingen, FRG.
The International Journal of Biological Markers
|April 1, 1990
Summary
Proficiency testing of the CA 125 assay revealed significant batch variations not detected by manufacturer controls. This highlights the need for robust quality assurance programs to ensure reliable laboratory test results.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Quality Assurance
Background:
- Reliable laboratory test results are crucial for patient diagnosis and management.
- Quality assurance programs are essential for validating the accuracy of diagnostic tests.
- The CA 125 assay is a significant biomarker, necessitating rigorous performance evaluation.
Purpose of the Study:
- To assess the performance and reliability of the CA 125 test system.
- To evaluate the effectiveness of proficiency testing in identifying assay variability.
- To establish a concentration-dependent performance profile for the CA 125 assay.
Main Methods:
- Proficiency testing conducted in five university laboratories over three years (1984-1987).
- Utilized five distinct reference materials for CA 125 analysis.
- Evaluated a total of 301 assays to establish performance profiles and precision.
Main Results:
- Intra-assay precision ranged from 4.8% to 11.5%; inter-assay precision ranged from 13.6% to 19.1%.
- Significant batch-dependent variations were observed, which were not flagged by manufacturer-supplied kit controls.
- No systematic drift was detected over the study period, indicating excellent stability of frozen reference materials.
Conclusions:
- Strict quality assurance programs, including proficiency testing, are vital for ensuring the reliability of CA 125 laboratory test results.
- Manufacturer kit controls may not adequately detect critical batch-to-batch variations in the CA 125 assay.
- The stability of properly stored reference materials is confirmed, supporting their use in quality control.