Related Experiment Video
Updated: May 20, 2026

Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
An acoustic sensor for monitoring airflow in pediatric tracheostomy patients
Thomas Ruscher1, Alexandrina Wicks Phd, Andre Muelenaer Md
1Virginia Tech.
Insights
Ultrasonic time-of-flight (TOF) offers a novel, non-invasive method for monitoring artificial airways. This technology accurately measures airflow, reducing false alarms and improving patient safety, especially for children.
Area of Science:
- Biomedical Engineering
- Medical Device Technology
- Respiratory Monitoring
Background:
- Patients with artificial airways face risks like airway obstruction from dislodged tubes or mucus.
- Young children are particularly vulnerable due to difficulties in self-care and the limitations of current indirect outpatient monitoring methods (e.g., thoracic impedance, pulse oximetry), which often yield false alarms.
- Caregiver desensitization to frequent false alarms is a documented factor in pediatric mortality.
Purpose of the Study:
- To evaluate the efficacy of ultrasonic time-of-flight (TOF) as a non-invasive monitoring technique for patients with tracheostomy tubes.
- To develop and test a prototype device for accurate measurement of respiratory parameters in patients with artificial airways.
Main Methods:
- Utilized ultrasonic time-of-flight (TOF), a technique adapted from industrial flow measurement, employing two transducers across a flow channel to detect acoustic signal transit time variations.
- Designed and constructed two generations of a prototype device, with the second generation featuring a smaller form factor and advanced electronics.
- Tested the prototype's ability to measure inspired volume non-invasively.
Main Results:
- The ultrasonic TOF technique demonstrated immunity to partial tube fouling (e.g., mucus, condensation) due to its measurement principle.
- The second-generation prototype successfully measured inspired volume with a root-mean-square error of less than 2% in initial trials.
- The technology provides a direct, non-invasive measurement, unlike indirect methods prone to false positives.
Conclusions:
- Ultrasonic time-of-flight (TOF) is a promising technology for accurate and reliable non-invasive monitoring of artificial airways.
- This technique has the potential to significantly improve patient safety by reducing false alarms and enabling better management of respiratory status, particularly in vulnerable pediatric populations.
- Further development could lead to improved clinical outcomes for patients requiring tracheostomy monitoring.
Abstract:
Without proper monitoring, patients with artificial airways in the trachea are at high risk for complications or death. Despite routine maintenance of the tube, dislodged or copious mucus can obstruct the airway. Young children ( 3yrs) have difficulty tending to their own tubes and are particularly vulnerable to blockages. They require external respiratory sensors. In a hospital environment, ventilators, end-tidal CO2 monitors, thermistors, and other auxiliary equipment provide sufficient monitoring of respiration. However, outpatient monitoring methods, such as thoracic impedance and pulse oximetry, are indirect and prone to false positives. Desensitization of caregivers to frequent false alarms has been cited in medical literature as a contributing factor in cases of child death. Ultrasonic time-of-flight (TOF) is a technique used in specialized industrial applications to non-invasively measure liquid and gas flow. Two transducers are oriented at a diagonal across a flow channel. Velocity measurement is accomplished by detecting slight variations in transit time of contra-propagating acoustic signals with a directional component parallel to air flow. Due to the symmetry of acoustic pathway between sensors, velocity measurements are immune to partial fouling in the tube from mucus, saliva, and condensation. A first generation proof of concept prototype was constructed to evaluate the ultrasonic TOF technique for medical tracheostomy monitoring. After successful performance, a second generation prototype was designed with a smaller form factor and more advanced electronics. This prototype was tested and found to measure inspired volume with a root-mean-square error < 2% during initial trials.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Tracheostomy Suctioning II: Procedure