Cardiac computed tomography angiography with automatic tube potential selection: effects on radiation dose and image
Brian B Ghoshhajra1, Leif-Christopher Engel, Mihály Károlyi
1Department of Radiology, Cardiac MR PET CT Program, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA. bghoshhajra@partners.org
Purpose:
Automatic exposure control (AEC) algorithms are widely available in coronary computed tomography angiography (CTA) and have been shown to reduce radiation doses by adjusting tube current to patient size. However, the effects of anthropometry-based automatic potential selection (APS) on image quality and radiation dose are unknown. We sought to investigate the effect of an APS algorithm on coronary CTA radiation dose and image quality.
Materials And Methods:
For this retrospective case-control study we selected 38 patients who had undergone coronary CTA for coronary artery assessment in whom tube potential and tube current were selected automatically by a combined automatic tube potential and tube current selection algorithm (APS-AEC) and compared them with 38 controls for whom tube voltage was selected according to standard body mass index (BMI) cutoffs and tube current was selected using automatic exposure control (BMI-AEC). Controls were matched for BMI, heart rate, heart rhythm, sex, acquisition mode, and indication for cardiac CTA. Image quality was assessed as contrast-to-noise ratio and signal-to-noise ratio in the proximal coronary arteries. Subjective reader assessment was also made. Total radiation dose (volume-weighted computed tomography dose index) was measured and compared between the 2 groups. In the study group, comparison was made with conventional BMI-guided prior protocols (site protocols and Society of Cardiovascular Computed Tomography recommendations) through disagreement analysis.
Results:
The APS-AEC cases received 29.8% lower overall radiation dose compared with controls (P=not significant). APS-AEC resulted in a significantly higher signal-to-noise ratio of the proximal coronary arteries (P<0.01) and contrast-to-noise ratio of the left main (P=0.01). In the study cases, the APS resulted in a change in tube potential versus site protocols and Society of Cardiovascular Computed Tomography recommendations in 45% (n=17) and 50% (n=19) of patients, respectively.
Conclusion:
Automated tube potential selection software resulted in significantly improved objective image quality versus standard BMI-based methods of tube potential selection, without increased radiation doses.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...