Feasibility study of automated overnight closed-loop glucose control under MD-logic artificial pancreas in patients
Revital Nimri1, Eran Atlas, Michal Ajzensztejn
1Jesse Z. and Sara Lea Shafer Institute for Endocrinology and Diabetes, The National Center for Childhood Diabetes, Schneider Children's Medical Center of Israel, 14 Kaplan St., Petah Tikva, Israel .
Background:
Artificial pancreas systems may offer a potential major impact on the normalization of metabolic control and preventing hypoglycemic events. This study aims to establish near-normal overnight glucose control and reduce the risk of nocturnal hypoglycemia using the MD-Logic Artificial Pancreas (MDLAP), an algorithm that was developed by our research group. This inpatient feasibility study is the first step towards implementing an overnight closed-loop MDLAP system at the patient's home.
Subjects And Methods:
Seven patients with type 1 diabetes (three adolescents and four adults; mean±SD age, 20.6±4.7 years; duration of diabetes, 9.6±2.6 years; body mass index, 24.3±3.9 kg/m(2); and glycated hemoglobin, 7.8±0.8%) participated in a total of 14 closed-loop overnight sessions. Each participant underwent two closed-loop inpatient sessions starting at dinner alone and at dinner following exercise. The closed-loop inpatient sessions were compared with data derived from nights spent at home with an open-loop system in a similar scenario to the study protocol.
Results:
The mean percentage of time spent in the near normal glucose range of 63-140 mg/dL was 83±16%, and the median (interquartile range) was 85% (78-92%) for the overnight closed-loop sessions compared with 34±31% and 27% (6-57%) in the homecare open-loop setting, respectively. During the overnight closed-loop sessions at dinner alone 92±9% of the sensor values ranged within target range, compared with 73±19% for the sessions following exercise (P=0.03). No hypoglycemic (<63 mg/dL) events occurred during the closed-loop sessions.
Conclusion:
Closed-loop insulin delivery under MDLAP is a feasible and safe solution to control overnight glycemia.
Related Concept Videos
Type I Diabetes I: Introduction
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Type I Diabetes II: Pathophysiology
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Type I Diabetes III: Clinical Manifestations