Related Experiment Video
Updated: May 19, 2026

How to Build a Laser Speckle Contrast Imaging (LSCI) System to Monitor Blood Flow
Published on: November 11, 2010
64×64 pixel smart sensor array for laser Doppler blood flow imaging
D He1, H C Nguyen, B R Hayes-Gill
1Faculty of Engineering, University of Nottingham, University Park, Nottingham, UK.
Abstract:
What is believed to be the first fully integrated two-dimensional complementary metal oxide semiconductor (CMOS) imaging array for laser Doppler blood flow imaging is demonstrated. The sensor has 64×64 pixels and includes both analog and digital on-chip processing electronics. This offers several potential advantages over commercial sensors as the processing is tailored to the signals of interest and the data bottleneck that exists between the sensor and processing electronics is overcome. To obtain a space efficient design over 64×64 pixels means that standard processing electronics used off-chip cannot be implemented. Images of both simulated blood flow responses and a blood flow occlusion test demonstrate the capability.