Proximal thoracic aortic diameter measurements at CT: repeatability and reproducibility according to measurement

Leslie E Quint1, Peter S Liu, Anna M Booher

  • 1Department of Radiology, University of Michigan Health System, Ann Arbor, MI 48109, USA. lequint@umich.edu

Insights

Variability in CT measurements of proximal thoracic aorta diameters was assessed. Both double-oblique short axis and semiautomatic centerline analysis showed low variability, but differences up to 5 mm occurred, with neither technique proving superior.

Area of Science:

  • Radiology
  • Cardiovascular Imaging
  • Medical Imaging Analysis

Background:

  • Accurate measurement of proximal thoracic aorta diameters is crucial for diagnosing and managing aortic diseases.
  • Computed tomography (CT) is a primary imaging modality for evaluating the aorta.
  • Standardized measurement techniques are essential for reliable interpretation of CT scans.

Purpose of the Study:

  • To compare the variability of proximal thoracic aortic diameter measurements using double-oblique short axis and semiautomatic centerline analysis techniques.
  • To assess intraobserver and interobserver variability for each measurement technique.
  • To evaluate the agreement between the two measurement techniques.

Main Methods:

  • Retrospective analysis of cardiac-gated thoracic aortic CT scans from 25 patients.
  • Measurements of aortic diameter at the annulus, sinuses, sinotubular junction, and ascending aorta.
  • Application of double-oblique short axis and semiautomatic centerline analysis techniques.
  • Statistical assessment using Wilcoxon signed rank test, Spearman's correlation, and Bland-Altman plots.

Main Results:

  • Mean intraobserver diameter differences were low for both techniques (double oblique: -0.3 to 0.6 mm; centerline: 0.2 to 2.3 mm).
  • Significant interobserver differences were observed for both methods.
  • Measurements were strongly correlated (r = 0.81-0.99) but consistently larger with centerline analysis.
  • 95% confidence intervals for differences between techniques ranged from ±1.8 to ±3.2 mm.

Conclusions:

  • Both double-oblique short axis and semiautomatic centerline analysis demonstrate generally low measurement variability for the proximal thoracic aorta.
  • Differences up to approximately 5 mm in diameter can occur within the 95% confidence interval between techniques.
  • Neither technique was definitively more reliable than the other, highlighting the need for careful interpretation and potential standardization.