Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors
Nicolas Nesseler1, Jean-Vincent Frénel, Yoann Launey
1Département d'Anesthésie-Réanimation 1, Inserm U991, Service de Réanimation Chirurgicale, Hôpital Pontchaillou, Université Rennes 1, 2 Rue Henri Le Guilloux, 35033 Rennes Cedex 9, France. nicolas.nesseler@chu-rennes.fr
Purpose:
Pulse oximetry (SpO(2)) measured at finger site via transmission mode may fail in situations of hypoperfusion. Forehead sensors using reflectance technology might be useful in these circumstances. We hypothesized that reflectance SpO(2) would be more accurate than finger SpO(2) in patients with severe shock.
Methods:
A prospective observational study was conducted in an intensive care unit of a university hospital of patients in shock who were treated with high norepinephrine and/or epinephrine doses (≥0.1 μg kg(-1) min(-1)). When blood gas determinations were requested, forehead SpO(2) and finger SpO(2) values were simultaneous recorded. Agreement between SpO(2) measurements with arterial saturation (SaO(2)), obtained by blood analysis with a co-oximeter, was assessed using the Bland-Altman method. The number of outliers, defined by the formula SaO(2) - SpO(2) > ±3 %, indicated the proportion of measurements considered to be clinically unacceptable.
Results:
Thirty-two patients were enrolled in the study. With the forehead sensor no reading failure occurred, and 140 paired data sets (forehead SpO(2) vs. SaO(2)) were obtained. Bias and precision were +1.0 and +2.5 %, respectively, and the limits of agreement ranged from -4.0 to 6.0 %. The finger sensor failed to give a value in four cases, thus providing 136 paired data sets (finger SpO(2) vs. SaO(2)) for analysis. Bias and precision were +1.4 and +4.8 %, respectively, and the limits of agreement ranged from -8.0 to 10.9 %. There were 21 (15 %) outliers for the forehead sensor and 43 (32 %) for the finger sensor (P < 0.001).
Conclusions:
Forehead SpO(2) measurements were more accurate than finger SpO(2) when compared with SaO(2) in critically ill patients requiring high-dose vasopressor therapy and should therefore be the preferred method considered.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.