Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Comparison of postoperative refractive outcomes: IOLMaster® versus immersion ultrasound
Woong-Joo Whang1, Byung-Ju Jung, Tae-Hoon Oh
1The Catholic Institute for Visual Science, The Catholic University of Korea, Seoul, Korea.
Background And Objective:
To compare the postoperative refractive outcomes between IOLMaster biometry (Carl Zeiss Meditec, Inc., Dublin, CA) and immersion ultrasound biometry for axial length measurements.
Patients And Methods:
Refractive outcomes in 354 eyes were compared using the IOLMaster and the immersion ultrasound biometry. Predicted refraction was determined using manual keratometry and the SRK-T formula with personalized A-constant.
Results:
The axial lengths measured using the IOLMaster and immersion ultrasound were 24.49 ± 2.11 and 24.46 ± 2.11 mm, respectively, and the difference was significant (P < .05). The mean errors were 0.000 ± 0.578 D with the IOLMaster, and 0.000 ± 0.599 D with the immersion ultrasound, but the difference was not significant. The mean absolute error was smaller with the IOLMaster than with immersion ultrasound (0.463 ± 0.341 vs 0.479 ± 0.359 D), but the difference was not significant.
Conclusion:
IOLMaster biometry yields highly accurate results in cataract surgery. However, if the IOLMaster is unavailable, immersion ultrasound biometry with personalized intraocular lens constants is an acceptable alternative.