[Comparison of quasi-continuous and continuous tunable diode laser absorption spectroscopy for gas detection]
Zhen-Hui Du1, Dong-Yu Gao, Ru-Bin Qi
1State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China. duzhenhui@tju.edu.cn
Abstract:
The theoretical analysis of the direct absorption spectroscopy, the continuous modulation spectroscopy and the quasi-continuous modulation spectroscopy was shown and the corresponding experiments were carried out in order to choose the adequate scheme of the laser modulation spectroscopy to satisfy different requirements of the detection. CO2 gas with different concentrations was detected under the same experimental conditions by using the three different modulation techniques with the same laser. Technical characteristics, signal features and detection limits were compared respectively. Results showed that the detection limit of the quasi-continuous modulation spectroscopy was approaching to that of the continuous modulation spectroscopy. However the linear distortion of the detection signal was obvious, because of the effects of laser energy intermittent and parasitic amplitude modulation on the line shape. Therefore the quasi-continuous modulation spectroscopy is not suitable for the pressure and flow measurements, which closely depend on the line shape. This work has provided reference for selections of the laser modulation spectroscopy.
Related Concept Videos
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Flame Photometry: Overview
UV–Vis Spectrometers
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.