Correlation between CT attenuation value and iodine concentration in vitro: discrepancy between gemstone spectral
Le Wang1, Bin Liu, Xing-Wang Wu
1Department of Radiology, the First Affiliated Hospital of AnHui Medical University, Hefei, Japan.
Introduction:
To assess the relation between CT attenuation value and iodine concentration in vitro, using gemstone spectral imaging (GSI) with single-source dual-energy CT and traditional polychromatic X-ray imaging (TPXI), respectively.
Methods:
A polypropylene phantom with eight test tubes in which iodine concentrations of solution were 0.4, 0.7, 2, 5, 10, 20, 30 and 50 mg/mL underwent GSI and traditional polychromatic X-ray scans (80, 100, 120 and 140 kV(p)), using single-source dual-energy spectral CT (Discovery CT750HD; GE Healthcare Technologies, Milwaukee, WI, USA) at the same tube speed of 0.8 s/rotation. All spectral imaging data were analysed with GSI viewer to obtain monochromatic images (50-140 keV, interval of 10 keV). Computed tomography attenuation values of iodine solution were measured with the same size of regions of interest and at the exact same level for both monochromatic and polychromatic images. The relation between CT attenuation value and iodine concentration was examined.
Results:
A linear correlation was found between CT attenuation value and iodine concentration for both monochromatic and polychromatic images. Moreover, the fitting coefficients for CT attenuation values and iodine concentrations were closer to one with GSI (r(2) = 0.99824-0.99996) than that with TPXI (r(2) = 0.99640-0.99736).
Conclusions:
Owing to the better correlation coefficients between CT attenuation value and iodine concentration, GSI may be a preferred method for quantitative measurement compared with TPXI.
More Related Videos
08:09Microfocus X-ray CT (microCT) Imaging of Actinia equina (Cnidaria), Harmothoe sp. (Annelida), and Xenoturbella japonica (Xenacoelomorpha)
Published on: August 6, 2019
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET