Maximizing the peak capacity using coupled columns packed with 2.6 μm core-shell particles operated at 1200 bar
Axel Vaast1, Jelle De Vos, Ken Broeckhoven
1Vrije Universiteit Brussel, Department of Chemical Engineering, Pleinlaan 2, B-1050 Brussels, Belgium.
Abstract:
We report on the peak capacity that can be produced by operating a state-of-the-art core-shell particle type (d(p)=2.6 μm) at its kinetic optimum at ultra-high pressures of 600 and 1200 bar. The column-length optimization needed to arrive at this kinetic optimum was realized using column coupling. Whereas the traditional operating mode (using a single 15 cm column operated at its optimum flow rate of 0.4 mL/min) offered a peak capacity of 162 in 10.8 min, a fully optimized train of 60 cm (4×15 cm) columns offered a peak capacity of 325 in 61 min when operated at 1200 bar. Even though the particles have a reputed low flow resistance and a relatively large size (>2 μm), it was found that the increase in performance that can be generated when switching from a fully optimized 600 bar operation to a fully optimized 1200 bar operation is significant (roughly 50% reduction of the analysis time for the same peak capacity and approximately a 20% increase in peak capacity if compared for the same analysis time). This has been quantified in a generic way using the kinetic plot method and is illustrated by showing the chromatograms corresponding to some of the data points of the kinetic plot curve.
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...