Related Experiment Video
Updated: May 19, 2026

Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Thermodilution and Fick cardiac outputs differ: impact on pulmonary hypertension evaluation
Wassim H Fares1, Sarah K Blanchard, George A Stouffer
1Yale University, School of Medicine, Department of Internal Medicine, Section of Pulmonary and Critical Care Medicine, Pulmonary Vascular Disease Program, New Haven, Connecticut 06510, USA. wassim.fares@yale.edu
Background:
The relationship between thermodilution and indirect Fick cardiac output determination methods has not been well described.
Objective:
To describe the relationship between these two cardiac output determination methods in patients evaluated for pulmonary hypertension and to highlight potential clinical implications.
Methods:
A retrospective review of charts of all adult patients who underwent a right heart catheterization (RHC) between January 1, 2007 and November 10, 2010, and participated in the pulmonary hypertension program of the pulmonary division at an academic institution was conducted. For validation, the charts of all patients who underwent RHC during the same period within the cardiology division were reviewed.
Results:
A total of 198 patients underwent 213 RHCs, 79 (40%) of whom had pulmonary arterial hypertension, were included. Forty-three per cent of patients had >20% difference between thermodilution and Fick. The average difference (thermodilution - Fick ±SD) was -0.39±2.03 L⁄min (n=213; P=0.006). There was no significant difference in bias or variability between thermodilution and Fick among patients with tricuspid regurgitant jet velocity (TRJ) of <3 m⁄s versus those with TRJ >3 m⁄s (-0.41±2.10 L⁄min versus -0.36±1.93 L⁄min, respectively; P=0.87). In a multivariable analysis, the thermodilution-Fick difference increased with age (P=0.001).
Discussion:
The presence of such discrepancy in 36% of patients evaluated for heart failure and⁄or heart transplant validated the results. In total, 37% of the 1315 procedures (213 performed by pulmonologists and 1102 performed by cardiologists) had a difference of >20% between thermodilution and Fick.
Conclusion:
Significant discrepancy exists between thermodilution and indirect Fick methods. This discrepancy potentially impacts pulmonary arterial hypertension prognostication and diagnosis, and is independent of TRJ.
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Mitral Stenosis I: Introduction
Cardiac Catheterization II: Right Heart Catheterization