Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Effect of jaw size in megavoltage CT on image quality and dose
Jae Hong Jung1, Kwang Hwan Cho, Yong Ho Kim
1Department of Radiation Oncology, College of Medicine, Soonchunhyang University Bucheon Hospital, Bucheon 1174, Korea.
Purpose:
Recently, the jaw size for the TomoTherapy Hi-Art II(®) (TomoTherapy Inc., Madison, WI) was reduced from 4 mm (J4) to 1 mm (J1) to improve the longitudinal (IEC-Y) resolution in megavoltage computed tomography (MVCT) images. This study evaluated the effect of jaw size on the image quality and dose, as well as the dose delivered to the lens of the eye, which is a highly radiosensitive tissue.
Methods:
MVCT image quality (image noise, uniformity, contrast linearity, high-contrast resolution, and full width at half-maximum) and multiple scan average dose (MSAD) were measured at different jaw sizes. A head phantom and photoluminescence glass dosimeters (PLDs) were used to measure the exposed lens dose (cGy). Different MVCT scan modes (pitch = 1, 2, and 3) and scan lengths (108 mm, 156 mm, and 204 mm) were applied in the MSAD and PLDs measurements.
Results:
The change in jaw size from J4 to J1 produced no change or only a slight improvement in image noise, uniformity, contrast linearity, and high-contrast resolution. However, the full-width at half-maximum reduced from approximately 7.2 at J4 to 4.5 mm at J1, which represents an enhancement in the longitudinal resolution. The MSAD at the center point changed from approximately 0.69-2.32 cGy (peripheral: 0.83-2.49 cGy) at J4 to 0.85-2.81 cGy (peripheral: 1.05-2.86 cGy) at J1. The measured lens dose increased from 0.92-3.36 cGy at J4 to 1.06-3.91 cGy at J1.
Conclusions:
The change in jaw size improved longitudinal resolution. The MVCT imaging dose of approximately 3.86 cGy, 1.92 cGy, and 1.22 cGy was delivered at a pitch of 1, 2, and 3, respectively, per fraction in the head and neck treatment plans. Therefore, allowance for an approximately 15% increase in lens dose over that with J4 should be provided with J1.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT