Effective Analysis of Human Exposure Conditions with Body-worn Dosimeters in the 2.4 GHz Band
Published on: May 2, 2018
SAR measurement due to mobile phone exposure in a simulated biological media
1Bioelectromagnetics Laboratory, School of Environmental Sciences, Jawaharlal Nehru University, New Delhi, India. jbehari@hotmail.com
Abstract:
The specific absorption rate (SAR) measurements are carried out for compliance testing of personal 3G Mobile phone. The accuracy of this experimental setup has been checked by comparing the SAR in 10 gm of simulated tissue and an arbitrary shaped box. This has been carried out using a 3G mobile Phone at 1718.5 MHz, in a medium simulating brain and muscle phantom. The SAR measurement system consists of a stepper motor to move a monopole E-field probe in two dimensions inside an arbitrary shaped box. The phantom is filled with appropriate frequency-specific fluids with measured electrical properties (dielectric constant and conductivity). That is close to the average for gray and white matters of the brain at the frequencies of interest (1718.5 MHz). Induced fields are measured using a specially designed monopole probe in its close vicinity. The probe is immersed in the phantom material. The measured data for induced fields are used to compute SAR values at various locations with respect to the mobile phone location. It is concluded that these SAR values are position dependent and well below the safety criteria prescribed for human exposure.