Related Experiment Video
Updated: May 19, 2026

Comparative Analysis of Human Growth Hormone in Serum Using SPRi, Nano-SPRi and ELISA Assays
Published on: January 7, 2016
Automated 22-kD growth hormone-specific assay without interference from Pegvisomant
Jenny Manolopoulou1, Younes Alami, Stephan Petersenn
1Endocrine Research Laboratories, Medizinische Klinik und Poliklinik IV, Ludwig Maximilians University, Munich, Germany.
A new growth hormone (GH) assay on the IDS-iSYS platform accurately measures GH levels, overcoming interference from Pegvisomant and GH binding protein (GHBP). This reliable assay meets consensus standards for GH measurement.
Area of Science:
- Endocrinology
- Assay Development
- Biochemical Analysis
Background:
- Significant variability exists in growth hormone (GH) assays due to calibration, antibody specificity, and interference.
- GH receptor antagonist Pegvisomant poses a challenge for existing GH assays.
- Consensus recommendations were established for GH assay standardization and evaluation.
Purpose of the Study:
- To develop and evaluate a new GH assay on the automated IDS-iSYS analyzer.
- To address interference issues encountered with Pegvisomant and GH binding protein (GHBP).
- To meet the criteria set forth by a recent consensus conference on GH assay standardization.
Main Methods:
- Developed a new GH assay using monoclonal antibodies, one specific for 22-kD GH and non-reactive to Pegvisomant.
- Assessed isoform specificity and GHBP interference, comparing with existing Siemens Immulite and Diasorin Liaison assays.
- Measured GH concentrations in healthy volunteers and acromegaly patients, and analyzed GH during Pegvisomant treatment.
Main Results:
- The IDS-iSYS GH assay demonstrated high analytical and functional sensitivity (0.01 μg/L and 0.04 μg/L).
- Achieved low intra-assay (2-4%) and inter-assay (5-7%) coefficients of variation.
- The assay showed specificity for 22-kD GH, was unaffected by GHBP, and importantly, was not impacted by Pegvisomant, unlike other assays.
Conclusions:
- The new IDS-iSYS GH assay adheres to recent consensus recommendations for standardization.
- It provides a reliable and accurate tool for measuring growth hormone levels.
- The assay effectively overcomes interference from Pegvisomant and GHBP, improving diagnostic utility.
More Related Videos
08:06Automated Acoustic Dispensing for the Serial Dilution of Peptide Agonists in Potency Determination Assays
Published on: November 10, 2016
08:28Validated Immunochemical Assay for Comprehensive Determination of the Human Epidermal Growth Factor Receptor 2 Released from and Bound to Cells
Published on: May 9, 2025