An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Relation between initial implant stability quotient and bone-implant contact percentage: an in vitro model study
Heng-Li Huang1, Ming-Tzu Tsai, Kuo-Chih Su
1Associate Professor, School of Dentistry, China Medical University, Taichung, Taiwan.
Objective:
This study evaluated the correlation between the initial stability of a dental implant, quantified as the implant stability quotient (ISQ), and the bone-implant contact percentage ( BIC%).
Study Design:
Dental implants were inserted into specimens comprising an artificial cortical shell representing cortical bone and foam bone representing cancellous bone with 4 elastic moduli. Each specimen with an implant was subjected to microcomputed tomography (micro-CT) scanning, from which the 2D and 3D BIC% values were calculated. The values of the Spearman correlation coefficient (r) were calculated between the ISQ and the 2D and 3D BIC% values.
Results:
Increasing the elasticity of the specimen enhanced the ISQ. There was statistically significant correlation between the ISQ and 3D BIC% values (r = 0.85, P < .0001). However, the correlation between ISQ and 2D BIC% values was not statistically significance (r = 0.42, P = .062).
Conclusions:
The initial implant stability was strongly positively correlated with the 3D BIC%.