Reocclusion caused by stent fracture implanted in the subclavian artery ostium: a case report
Norihiko Shinozaki1, Ralf Langhoff, Karl-Ludwig Schulte
1Department of Cardiology, Tokai University School of Medicine, 143 Shimokasuya, Isehara, Kanagawa, Japan. shinori3@tokai-u.jp
Abstract:
A 47 year-old male, a woodcutter by profession, complained of severe pain in the left arm during physical stress. Eighteen months ago, a balloon-expandable stainless steel stent had been implanted in the left subclavian artery ostium. The angiography revealed a stent fracture at the origin of the left subclavian artery, this condition led to a re-obstruction of the stent. After the insertion of a self-expandable Nitinol stent, the side differences of arterial pressure and the symptoms disappeared. This is the first reported case of a stent fracture inserted at the origin of the subclavian artery without severe calcification.