Related Experiment Video
Updated: May 19, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
"Add-on pacing lead": an effective and safe alternative to lead replacement in ICD pacing failure
Sudeep Kumar1, Nagaraja Moorthy, Aditya Kapoor
1Department of Cardiology, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow, India. sudeep@sgpgi.ac.in
Abstract:
Despite recent advances in implantable cardioverter defibrillator (ICD) technology, the long-term reliability of ICD leads remains a significant problem. Lead failures constitute a major risk for patients with an implantable cardioverter defibrillator. There is no clear consensus on treatment strategy of ICD lead failure and decision should be individualized. We report a pacing-dependent elderly male with ICD lead pacing failure secondary to insulation break resulting in recurrent syncope. We emphasize the technique of "Add-on pacing lead implantation" could be an effective and alternative to ICD lead replacement.