Clinical performance of 2 dedicated PET scanners for breast imaging: initial evaluation
Mami Iima1, Yuji Nakamoto, Shotaro Kanao
1Department of Diagnostic Imaging and Nuclear Medicine, Kyoto University Graduate School of Medicine, Kyoto, Japan.
Unlabelled:
The purpose of this study was to investigate the diagnostic performance of 2 newly developed dedicated breast PET scanners in patients with known or suspected breast cancer.
Methods:
Two types of scanner were evaluated, an O-shaped scanner and a C-shaped scanner. The O scanner was designed for imaging patients who were prone, and the C scanner was designed for those patients positioned leaning forward. Sixty-nine women with known or suspected breast carcinoma (80 lesions: 72 invasive carcinomas, 4 noninvasive carcinomas [ductal carcinoma in situ, or DCIS], 1 case of adenomatous ductal hyperplasia, and 3 benign lesions) were enrolled in this study. All patients underwent a conventional whole-body PET/CT scan, followed by breast scanning using both dedicated devices. The diagnostic performance of each scanner was assessed.
Results:
The maximal diameter of invasive tumors ranged from 4 to 112 mm, with an average of 26 mm. With the O scanner, 62 of 76 malignant lesions (including 3 DCIS) were detected, 5 lesions were not detected, and the remaining 9 lesions were outside the field of view. With the C scanner, 63 of 76 malignant lesions (including 2 DCIS) were detected, 7 lesions were not detected, and the remaining 6 lesions were outside the field of view. The lesion-based sensitivities of the O and C scanners were 82% (62/76) and 83% (63/76), respectively; sensitivities excluding lesions outside the field of view were 93% (62/67) and 90% (63/70), respectively. The sensitivity of conventional PET/CT was 92% (70/76). All lesions outside the field of view were close to the chest wall. The breast-based specificities of the O, C, and conventional scanners were 98% (48/49), 98% (56/57), and 100% (70/70), respectively.
Conclusion:
Our preliminary study indicates that both dedicated breast PET scanners are clinically feasible and yield reasonably high sensitivity. More detailed information was obtained with these scanners than with the conventional scanner.
More Related Videos
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies II: Ultrasonography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...