Study on probe current dependence of the intensity distribution in annular dark field images
Suhyun Kim1, Yoshifumi Oshima, Yasumasa Tanishiro
1Department of Physics, Tokyo Institute of Technology, 2-12-1, 1-24 Oh-okayama, Tokyo 152-8551, Japan. u98kim@surface.phys.titech.ac.jp
Abstract:
The intensity profile of atomic columns in annular dark field (ADF) images was quantitatively investigated for a silicon (001) crystal using a cold field emission source in an aberration corrected scanning transmission electron microscope. The intensity distribution at the atomic column in the annular dark field image was blurred by increasing the probe current from 10 to 40 pA, which was quantitatively well fit by a simulated distribution convolved by Gaussian envelopes with area proportional to the probe current. The blur of the ADF images was primarily determined by the size of the cold field emission source.
More Related Videos
Related Concept Videos
Overview of Microscopy Techniques
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...