Influence of oxygenator type on the prevalence and extent of microembolic retinal ischemia during cardiopulmonary

C I Blauth1, P L Smith, J V Arnold

  • 1Department of Surgery, RPMS, Hammersmith Hospital, London, England.

Insights

Cardiopulmonary bypass can cause retinal microemboli. Membrane oxygenators significantly reduce these microembolic perfusion defects compared to bubble oxygenators, offering better retinal protection.

Area of Science:

  • Cardiovascular Surgery
  • Ophthalmology
  • Biomedical Engineering

Background:

  • Cardiopulmonary bypass (CPB) is associated with microembolic events affecting the retina.
  • Quantifying retinal ischemia during CPB is crucial for understanding neurological complications.

Purpose of the Study:

  • To compare the incidence and extent of retinal microembolic ischemia between bubble and membrane oxygenators during CPB.
  • To evaluate the efficacy of digital image analysis in quantifying retinal microemboli.

Main Methods:

  • A prospective study of 64 patients undergoing elective coronary operations.
  • Retinal fluorescein angiography with digital image analysis to quantify microembolic perfusion defects.
  • Comparison between bubble oxygenators (n=30) and flat sheet membrane oxygenators (n=34).

Main Results:

  • All patients (100%) using bubble oxygenators showed retinal perfusion defects.
  • Only 44% of patients using membrane oxygenators had defects, a statistically significant reduction (p < 0.001).
  • Membrane oxygenators resulted in fewer lesions and smaller ischemic areas compared to bubble oxygenators.

Conclusions:

  • Digital image analysis of retinal fluorescein angiograms is a viable method for quantifying CPB-induced microembolic ischemia.
  • Flat sheet membrane oxygenation provides superior protection against retinal microemboli compared to bubble oxygenation during CPB.