Self-expandable stainless steel braided endoprosthesis for biliary strictures
A Gillams1, R Dick, J S Dooley
1Department of Radiology, Royal Free Hospital, London, England.
Radiology
|January 1, 1990
Summary
A novel stainless steel biliary stent was evaluated in 45 patients. While effective in delivery and non-migratory, it did not reduce recurrent jaundice compared to plastic stents.
Area of Science:
- Gastroenterology
- Interventional Radiology
- Biomedical Engineering
Background:
- Symptomatic biliary strictures, both malignant and benign, necessitate effective stent placement.
- Traditional biliary stents face challenges with diameter, migration, and recurrent complications.
- A new stainless steel woven mesh stent offers a large-diameter, percutaneously delivered option.
Purpose of the Study:
- To evaluate the safety and efficacy of a novel stainless steel woven mesh biliary stent.
- To assess complication rates, including mortality, early and late complications, and stent migration.
- To compare the frequency of recurrent jaundice with this new stent against historical data from plastic stents.
Main Methods:
- A prospective study involving 45 patients with symptomatic biliary strictures (40 malignant, 5 benign).
- Insertion of a compressed stainless steel woven mesh stent via a 7- or 9-F delivery catheter.
- Monitoring for 30-day mortality, early (<72 hours) and late complications, recurrent jaundice, and stent migration.
Main Results:
- The 30-day mortality rate was 7%, with an early complication rate of 16%.
- Recurrent jaundice occurred in 42% of patients, and late sepsis without jaundice in 11%.
- No instances of stent migration were observed.
Conclusions:
- The new large-diameter stainless steel biliary stent can be percutaneously inserted without increased risk of complications.
- The stent's non-migratory design is a significant advantage.
- The frequency of recurrent jaundice was similar to plastic stents, potentially due to patient selection and complexity of strictures.