An independent, objective calibration check for the reduced oxygen breathing device
David L Still1, Leonard A Temme
1Vision Sciences Branch, Sensory Research Division, U. S. Army Aeromedical Research Laboratory, Fort Rucker, AL 36362, USA. david.l.still@us.army.mil
Introduction:
Normobaric hypoxia, which does not entail an altitude chamber, but reduces the fraction of inspired oxygen (02) by diluting air with nitrogen, is finding increased use. The reduced oxygen breathing device (ROBD-2) is one of several commercial devices for generating such normobaric hypoxia. Reported here are results of a procedure to check the calibration of the ROBD-2 using methods that may be readily available in physiology and psychophysiology facilities.
Methods:
The %O2 output by the ROBD-2 was measured concurrently in two ways for altitudes from mean sea level (MSL) to 34,000 ft above MSL at 2000-ft intervals five times over 2 d. One measurement method used was the one built into the ROBD-2, which reports the %O2 the device is delivering at the selected target altitude. The other method diverted a sample of the ROBD-2's output gas to the paramagnetic O2, sensor of a metabolic measuring system via its sampling line. The %O2, measured with the two techniques was compared using Bland-Altman statistical procedures.
Results:
The two measurement methods produced %O2 readings differing by no more than 0.18% O2 from MSL to 34,000 ft (from 20.95 to about 4.40% oxygen, respectively), the full operating range of the device. Calculating altitude from the measured %O2 showed the ROBD-2 operated within its design error margins over its whole operational range.
Discussion:
The purpose for which the ROBD-2 is used should determine whether the reliability of its output in normobaric equivalent altitude is adequate. Differences between devices and device stability over time and with use have yet to be assessed. Our assessment does not address the accuracy of the algorithm the ROBD-2 uses to provide an equivalent target altitude under normobaric conditions.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...