Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Home-based overnight transcutaneous capnography/pulse oximetry for diagnosing nocturnal hypoventilation associated
Kristy A Bauman1, Armando Kurili, Shelley L Schmidt
1Pulmonary and Critical Care Medicine Division, Department of Internal Medicine, University of Michigan Health System, Ann Arbor, MI. krisbaum@umich.edu
Objective:
To determine the utility of home-based, unsupervised transcutaneous partial pressure of carbon dioxide (tc-Pco(2)) monitoring/oxygen saturation by pulse oximetry (Spo(2)) for detecting nocturnal hypoventilation (NH) in individuals with neuromuscular disorders.
Design:
Retrospective case series analyzed consecutively.
Setting:
Multidisciplinary neuromuscular respiratory failure (NMRF) clinic at an academic institution.
Participants:
Subjects (N=35, 68.6% men; mean age, 46.9y) with spinal cord injury (45.7%) or other neuromuscular disorders underwent overnight tests with tc-Pco(2)/Spo(2) monitoring. Fifteen (42.9%) were using nocturnal ventilatory support, either bilevel positive airway pressure (BiPAP) or tracheostomy ventilation (TV).
Interventions:
A respiratory therapist brought a calibrated tc-Pco(2)/Spo(2) monitor to the patient's home and provided instructions for data collection during the subject's normal sleep period. Forced vital capacity (FVC), body mass index (BMI), and exhaled end-tidal Pco(2) (ET-Pco(2)) were recorded at a clinic visit before monitoring.
Main Outcome Measures:
Detection of NH (tc-Pco(2) ≥50mmHg for ≥5% of monitoring time). Data were also analyzed to determine whether nocturnal oxygen desaturation (Spo(2) ≤88% for ≥5% of monitoring time), FVC, BMI, or daytime ET-Pco(2) could predict the presence of NH.
Results:
NH was detected in 18 subjects (51.4%), including 53.3% of those using BiPAP or TV. NH was detected in 43.8% of ventilator-independent subjects with normal daytime ET-Pco(2) (present for 49.4%±31.5% [mean ± SD] of the study period), and in 75% of subjects with an elevated daytime ET-Pco(2) (present for 92.3%±8.7% of the study period). Oxygen desaturation, BMI, and FVC were poor predictors of NH. Only 3 attempted monitoring studies failed to produce acceptable results.
Conclusions:
Home-based, unsupervised monitoring with tc-Pco(2)/Spo(2) is a useful method for diagnosing NH in NMRF.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.