Related Experiment Video
Updated: May 18, 2026

Using a 1064-nm Picosecond Neodymium-Doped Yttrium Aluminum Garnet Laser for Periorbital Hyperpigmentation
Published on: May 23, 2025
Utility of large spot binocular indirect laser delivery for peripheral photocoagulation therapy in children
Saranya C Balasubramaniam1, Brian G Mohney, Genie M Bang
1Department of Ophthalmology, Mayo Clinic and Mayo Foundation, Rochester, Minnesota 55905, USA.
Abstract:
The purpose of this article is to demonstrate the utility of the large spot size (LSS) setting using a binocular laser indirect delivery system for peripheral ablation in children. One patient with bilateral retinopathy of prematurity received photocoagulation with standard spot size burns placed adjacently to LSS burns. Using a pixel analysis program called Image J on the Retcam picture, the areas of each retinal spot size were determined in units of pixels, giving a standard spot range of 805 to 1294 pixels and LSS range of 1699 to 2311 pixels. Additionally, fluence was calculated using theoretical retinal areas produced by each spot size: the standard spot setting was 462 mJ/mm2 and the LSS setting was 104 mJ/mm2. For eyes with retinopathy of prematurity, our study shows that LSS laser indirect delivery halves the number of spots required for treatment and reduces fluence by almost one-quarter, producing more uniform spots.