Spectral analysis of bowel sounds in intestinal obstruction using an electronic stethoscope
Siok Siong Ching1, Yih Kai Tan
1Department of Surgery, Changi General Hospital, Singapore 529889, Singapore. ching_ss@hotmail.com
Aim:
To determine the value of bowel sounds analysis using an electronic stethoscope to support a clinical diagnosis of intestinal obstruction.
Methods:
Subjects were patients who presented with a diagnosis of possible intestinal obstruction based on symptoms, signs, and radiological findings. A 3M™ Littmann(®) Model 4100 electronic stethoscope was used in this study. With the patients lying supine, six 8-second recordings of bowel sounds were taken from each patient from the lower abdomen. The recordings were analysed for sound duration, sound-to-sound interval, dominant frequency, and peak frequency. Clinical and radiological data were reviewed and the patients were classified as having either acute, subacute, or no bowel obstruction. Comparison of bowel sound characteristics was made between these subgroups of patients. In the presence of an obstruction, the site of obstruction was identified and bowel calibre was also measured to correlate with bowel sounds.
Results:
A total of 71 patients were studied during the period July 2009 to January 2011. Forty patients had acute bowel obstruction (27 small bowel obstruction and 13 large bowel obstruction), 11 had subacute bowel obstruction (eight in the small bowel and three in large bowel) and 20 had no bowel obstruction (diagnoses of other conditions were made). Twenty-five patients received surgical intervention (35.2%) during the same admission for acute abdominal conditions. A total of 426 recordings were made and 420 recordings were used for analysis. There was no significant difference in sound-to-sound interval, dominant frequency, and peak frequency among patients with acute bowel obstruction, subacute bowel obstruction, and no bowel obstruction. In acute large bowel obstruction, the sound duration was significantly longer (median 0.81 s vs 0.55 s, P = 0.021) and the dominant frequency was significantly higher (median 440 Hz vs 288 Hz, P = 0.003) when compared to acute small bowel obstruction. No significant difference was seen between acute large bowel obstruction and large bowel pseudo-obstruction. For patients with small bowel obstruction, the sound-to-sound interval was significantly longer in those who subsequently underwent surgery compared with those treated non-operatively (median 1.29 s vs 0.63 s, P < 0.001). There was no correlation between bowel calibre and bowel sound characteristics in both acute small bowel obstruction and acute large bowel obstruction.
Conclusion:
Auscultation of bowel sounds is non-specific for diagnosing bowel obstruction. Differences in sound characteristics between large bowel and small bowel obstruction may help determine the likely site of obstruction.
Related Concept Videos
Assessment of the Abdomen I: Inspection and Auscultation
The abdominal examination is a cornerstone of clinical medicine, serving as a critical tool in diagnosing various gastrointestinal (GI) diseases. It involves a systematic approach that includes inspection and auscultation, each with distinct yet complementary roles in assessing the abdomen. This article will delve into these two primary methods healthcare professionals use to examine the abdomen.
Inspection of the Abdomen
The first step in any abdominal examination is inspection.
Intestinal Obstruction II: Pathophysiology
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Intestinal Obstruction I: Introduction
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.