Purpose:
Percutaneous spine procedures may occasionally be difficult and subject to complications. Navigation using a dynamic reference base (DRB) may ease the procedure. Yet, besides other shortcomings, its fixation demands additional incisions and thereby defies the percutaneous character of the procedure.
Methods:
A new concept of atraumatic referencing was invented including a special epiDRB. The accuracy of navigated needle placement in soft tissue and bone was experimentally scrutinised. Axial and pin-point deviations from the planned trajectory were investigated with a CT-based 3D computer system. Clinical evaluation in a series of ten patients was also done.
Results:
The new epiDRB proved convenient and reliable. Its fixation to the skin with adhesive foil provided a stable reference for navigation that improves the workflow of percutaneous interventions, reduces radiation exposure and helps avoid complications.
Conclusions:
Percutaneous spine interventions can be safely and accurately navigated using epiDRB with minimal trauma or radiation exposure and without additional skin incisions.