Published on: December 13, 2017
Rapid and sensitive cardiac troponin I immunoassay based on fluorescent europium(III)-chelate-dyed nanoparticles
Marja-Leena Järvenpää1, Katri Kuningas, Ilari Niemi
1Department of Biotechnology, University of Turku, Turku, Finland. mljarv@utu.fi.
Background:
Cardiac troponins are the preferred and recommended biomarkers of myocardial infarction. Unfortunately, most of the current commercial assays do not meet the guideline recommendations for sensitivity and low-end precision. Therefore, improvements in their analytical performance are still needed.
Methods:
Cardiac troponin I (cTnI) immunoassay was developed. The assay utilized a monoclonal antibody and a F(ab')(2) antibody fragment immobilized onto the microtiter wells for capturing, and a monoclonal antibody covalently conjugated to fluorescent europium(III)-chelate-dyed nanoparticles for detecting. Following a 15-min incubation of the sample and nanoparticle-bioconjugates in the capture wells, cTnI was quantified directly from the washed well surface by time-resolved fluorometry.
Results:
The limits of detection and quantification were 0.0020 μg/l and 0.012 μg/l, respectively. The response was linear in the measured range of 0.003-9.6 μg/l. The within-run imprecisions were 9.8, 5.1, 7.7 and 5.4%, and the total imprecisions were 13.1, 10.4, 9.0 and 8.7% at cTnI levels of 0.007, 0.051, 0.52 and 2.62 μg/l, respectively. Plasma recoveries of added cTnI were 72-119%. Regression analysis with Innotrac Aio! 2nd generation cTnI assay yielded a slope (95% confidence intervals) of 1.197 (1.141 to 1.253) and y-intercept of 0.216 (-0.128 to 0.561)μg/l (S(yx) = 2.176 μg/l, n = 212, r = 0.945).
Conclusions:
The developed immunoassay based on europium(III)-chelate-dyed nanoparticle label allows rapid and sensitive measurement of cTnI.
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.