Performance comparison of single-view digital breast tomosynthesis plus single-view digital mammography with two-view
Gisella Gennaro1, R Edward Hendrick, Patricia Ruppel
1Veneto Institute of Oncology (IOV)-IRCCS, via Gattamelata 64, 35128 Padua, Italy. gisella.gennaro@ioveneto.it
Objective:
To determine the performance of combined single-view mediolateral oblique (MLO) digital breast tomosynthesis (DBT) plus single-view cranio-caudal (CC) mammography (MX) compared with that of standard two-view digital mammography.
Methods:
A multi-reader multi-case (MRMC) receiver-operating characteristic (ROC) study was conducted, involving six breast radiologists. Two hundred fifty patients underwent bilateral MX and DBT imaging. MX and DBT images with the adjunct of the CC-MX view from 469 breasts were evaluated and rated independently by six readers. Differences in mean areas under the ROC curves (AUCs), mean sensitivity and mean specificity were analysed by analysis of variance (ANOVA) to assess clinical performance.
Results:
The combined technique was found to be non-inferior to standard two-view mammography (MX((CC+MLO))) in mean AUC (difference: +0.021;95 % LCL = -0.011), but was not statistically significant for superiority (P = 0.197). The combined technique had equivalent sensitivity to standard mammography (76.2 % vs. 72.8 %, P = 0.269) and equivalent specificity (84.9 % vs. 83.0 %, P = 0.130). Specificity for benign lesions was significantly higher with the combination of techniques versus mammography (45.6 % vs. 36.8 %, P = 0.002).
Conclusion:
In this enriched study population, the combination of single-view MLO tomosynthesis plus single-view CC mammography was non-inferior to that of standard two-view digital mammography in terms of ROC curve area, sensitivity and specificity.