Evaluation of Masimo signal extraction technology pulse oximetry in anaesthetized pregnant sheep
Christopher T Quinn1, Anthea L Raisis, Gabrielle C Musk
1Division of Veterinary and Biomedical Sciences, Murdoch University, Perth, WA, Australia. cquinn@csu.edu.au
Objective:
Evaluation of the accuracy of Masimo signal extraction technology (SET) pulse oximetry in anaesthetized late gestational pregnant sheep.
Study Design:
Prospective experimental study.
Animals:
Seventeen pregnant Merino ewes.
Methods:
Animals included in study were late gestation ewes undergoing general anaesthesia for Caesarean delivery or foetal surgery in a medical research laboratory. Masimo Radical-7 pulse oximetry (SpO(2) ) measurements were compared to co-oximetry (SaO(2) ) measurements from arterial blood gas analyses. The failure rate of the pulse oximeter was calculated. Accuracy was assessed by Bland & Altman's (2007) limits of agreement method. The effect of mean arterial blood pressure (MAP), perfusion index (PI) and haemoglobin (Hb) concentration on accuracy were assessed by regression analysis.
Results:
Forty arterial blood samples paired with SpO(2) and blood pressure measurements were obtained. SpO(2) ranged from 42 to 99% and SaO(2) from 43.7 to 99.9%. MAP ranged from 24 to 82 mmHg, PI from 0.1 to 1.56 and Hb concentration from 71 to 114 g L(-1) . Masimo pulse oximetry measurements tended to underestimate oxyhaemoglobin saturation compared to co-oximetry with a bias (mean difference) of -2% and precision (standard deviation of the differences) of 6%. Accuracy appeared to decrease when SpO(2) was <75%, however numbers were too small for statistical comparisons. Hb concentration and PI had no significant effect on accuracy, whereas MAP was negatively correlated with SpO(2) bias.
Conclusions And Clinical Relevance:
Masimo SET pulse oximetry can provide reliable and continuous monitoring of arterial oxyhaemoglobin saturation in anaesthetized pregnant sheep during clinically relevant levels of cardiopulmonary dysfunction. Further work is needed to assess pulse oximeter function during extreme hypotension and hypoxaemia.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.