Three commercial polyclonal immunoassays for cyclosporine in whole blood compared: 1. Results with patients'

M J Strassman1, G L Lensmeyer, D A Wiebe

  • 1Department of Pathology and Laboratory Medicine, University of Wisconsin Hospital & Clinics, Madison 53792.

Clinical Chemistry
|January 1, 1990
PubMed

Insights

Commercial cyclosporine (CsA) immunoassays show significant variability in whole-blood measurements. Patient results are not interchangeable between different assay methods, impacting clinical interpretation.

Area of Science:

  • Clinical Chemistry
  • Pharmacology
  • Immunology

Background:

  • Cyclosporine (CsA) is a critical immunosuppressant medication used in organ transplantation.
  • Accurate monitoring of CsA whole-blood concentrations is essential for therapeutic efficacy and preventing toxicity.
  • Several commercially available immunoassays exist for CsA measurement, but their performance characteristics may vary.

Purpose of the Study:

  • To evaluate and compare the performance of three commercially available polyclonal immunoassays for apparent cyclosporine (CsA) in whole blood.
  • To assess the interchangeability of CsA results obtained from different immunoassay methods.
  • To investigate the impact of CsA metabolites on assay accuracy and inter-method variability.

Main Methods:

  • Analyzed 120 whole-blood specimens from transplant recipients using Abbott TDx (fluorescent polarization immunoassay), Sandoz RIA, and Incstar Cyclo-Trac RIA.
  • Determined mean coefficients of variation (CVs), analytical recovery, and linearity for each assay.
  • Quantified parent CsA concentrations using High-Performance Liquid Chromatography (HPLC) for comparison.
  • Assayed serial specimens from six patients to track day-to-day fluctuations and inter-method agreement.

Main Results:

  • Mean CVs varied significantly across the assays: 3.8% (Abbott), 9.3% (Sandoz), and 24.3% (Incstar).
  • Analytical recovery was high for lower concentrations but linearity was compromised at higher levels (>1000-1500 µg/L).
  • Methods showed dissimilar but parallel results, with occasional deviations from established trends, partly due to cross-reactivity with CsA metabolites.
  • HPLC analysis confirmed discrepancies among immunoassay results.

Conclusions:

  • Commercially available polyclonal immunoassays for apparent cyclosporine demonstrate significant performance differences.
  • Interchangeability of patient results between these methods is not reliable due to assay biases and cross-reactivity with metabolites.
  • Clinical decisions based on cyclosporine levels should consider the specific assay method used.