Optimization of a reversed-phase-high-performance thin-layer chromatography method for the separation of isoniazid,
D H Shewiyo1, E Kaale, P G Risha
1Directorate of Laboratory Services, Tanzania Food and Drugs Authority, P.O. Box 77150, Dar es Salaam, Tanzania.
Abstract:
This paper presents the development of a new RP-HPTLC method for the separation of pyrazinamide, isoniazid, rifampicin and ethambutol in a four fixed-dose combination (4 FDC) tablet formulation. It is a single method with two steps in which after plate development pyrazinamide, isoniazid and rifampicin are detected at an UV wavelength of 280 nm. Then ethambutol is derivatized and detected at a VIS wavelength of 450 nm. Methanol, ethanol and propan-1-ol were evaluated modifiers to form alcohol-water mobile phases. Systematic optimization of the composition of each alcohol in the mobile phase was carried out using the window diagramming concept to obtain the best separation. Examination of the Rf distribution of the separated compounds showed that separation of the compounds with the mobile phase containing ethanol at the optimal fraction was almost situated within the optimal Rf-values region of 0.20-0.80. Therefore, ethanol was selected as organic modifier and the optimal mobile phase composition was found to be ethanol, water, glacial acetic acid (>99% acetic acid) and 37% ammonia solution (70/30/5/1, v/v/v/v). The method is new, quick and cheap compared to the actual method in the International Pharmacopoeia for the assay of the 4 FDC tablets, which involves the use of two separate HPLC methods.
Related Concept Videos
Pulmonary Tuberculosis V
Latent tuberculosis infection occurs when TB bacteria are present in a person's body, but are not causing illness or symptoms. It is not contagious, and preventive treatment is crucial to avoid the progression...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...