Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation
W Karlen1, K Kobayashi, J M Ansermino
1Electrical & Computer Engineering in Medicine Group, Department of Electrical & Computer Engineering, The University of British Columbia, 2332 Main Mall, Vancouver, BC, V6T 1Z4, Canada. walter.karlen@ieee.org
Abstract:
Pulse oximeters are monitors that noninvasively measure heart rate and blood oxygen saturation (SpO2). Unfortunately, pulse oximetry is prone to artifacts which negatively impact the accuracy of the measurement and can cause a significant number of false alarms. We have developed an algorithm to segment pulse oximetry signals into pulses and estimate the signal quality in real time. The algorithm iteratively calculates a signal quality index (SQI) ranging from 0 to 100. In the presence of artifacts and irregular signal morphology, the algorithm outputs a low SQI number. The pulse segmentation algorithm uses the derivative of the signal to find pulse slopes and an adaptive set of repeated Gaussian filters to select the correct slopes. Cross-correlation of consecutive pulse segments is used to estimate signal quality. Experimental results using two different benchmark data sets showed a good pulse detection rate with a sensitivity of 96.21% and a positive predictive value of 99.22%, which was equivalent to the available reference algorithm. The novel SQI algorithm was effective and produced significantly lower SQI values in the presence of artifacts compared to SQI values during clean signals. The SQI algorithm may help to guide untrained pulse oximeter users and also help in the design of advanced algorithms for generating smart alarms.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure