Update on the need for a permanent pacemaker after transcatheter aortic valve implantation using the CoreValve®
Didier Tchetche1, Thomas Modine, Bruno Farah
1Groupe CardioVasculaire Interventionnel, Clinique Pasteur, Toulouse, France. d.tchetche@clinique-pasteur.com
Aims:
High rates of permanent pacemaker (PPM) implantation are reported after transcatheter aortic valve implantation (TAVI) using the Medtronic CoreValve® system. The Accutrak™ catheter is designed to allow a more predictable landing zone. Little is known about the real clinical impact of this catheter. The aims of this paper were to describe the potential impact of the Accutrak™ catheter on the accuracy of positioning a 26 or 29 mm CoreValve® across the aortic annulus and its impact on the need for a pacemaker.
Methods And Results:
A total of 134 patients were treated with the CoreValve® Accutrak™ system at two French centres (Lille and Toulouse). Mean age was 82.4 ± 4.7 years; logistic EuroSCORE was 24.3 ± 9.5%. Procedural success rate was 99.2%; mean depth of implantation was 4.9 mm. A final position between 0 and 6 mm was achieved in 85.8% of the patients. All-cause mortality at 30 days was 6%. The PPM implantation rate was 10.6%. Due to a limited number of events, we could not identify any predictor of need for a PPM: pre-existing right bundle branch block (RBBB) (OR 2.72 [0.63-11.87], p=ns), use of a 29 mm prosthesis (OR 2.73 [0.33-22.90], p=ns) and left ventricular septal hypertrophy (OR 2.63 [0.08-83.32], p=ns).
Conclusions:
In this cohort of patients treated with the CoreValve® Accutrak™ system, the incidence of permanent pacemaker implantation was low, which may be a consequence of an average small implantation depth. The Accutrak™ catheter seems to be helpful in achieving higher and more predictable implants. Operators could standardise their technique to place the CoreValve® prostheses less than 6 mm below the aortic annulus.
Related Concept Videos
Aortic Regurgitation III: Medical Management
Mitral Stenosis III: Medical Management
Cardiac Catheterization IV: Nursing Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Regurgitation III: Medical Management
Mitral Valve Prolapse II: Assessment and Management