Structural valve deterioration in a starr-edwards mitral caged-disk valve prosthesis
Shigeaki Aoyagi1, Kei-Ichiro Tayama, Teiji Okazaki
1Department of Cardiovascular Surgery, Munakata Suikokai General Hospital, Fukutsu, Japan. shigeakiaoyagi@gmail.com
The Starr-Edwards (SE) mitral caged-disk valve, model 6520, shows rare structural deterioration. Disk fracture and wear were observed in valves implanted over 36 years, suggesting careful follow-up or elective replacement after 20-30 years.
Area of Science:
- Cardiovascular Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- The long-term durability of the Starr-Edwards (SE) mitral caged-disk valve, model 6520, remains incompletely understood.
- Structural valve deterioration (SVD) is considered rare for this specific mechanical heart valve model.
Purpose of the Study:
- To investigate the long-term structural integrity and potential deterioration of the SE mitral caged-disk valve.
- To determine the in-vivo performance and failure modes of SE mitral valves after extended implantation periods.
Main Methods:
- Analysis of explanted SE mitral disk valves (model 6520) from 7 patients.
- Macroscopic examination of valves removed between 23 and 40 years post-implantation.
- Assessment of structural abnormalities, disk wear, and disk fracture.
Main Results:
- No structural abnormalities were found in valves explanted before 26 years.
- Localized disk wear was observed in valves implanted for over 36 years.
- Disk fracture and multiple cracks were detected in valves implanted for 36 and 40 years.
Conclusions:
- Significant structural deterioration, including disk fracture and wear, can occur in SE mitral caged-disk valves after more than 36 years.
- Mechanical mitral valves implanted for over 20 years warrant close monitoring.
- Elective replacement with modern prosthetic valves is recommended for SE mitral valves implanted for over 30 years.