Comparison of the Icare tonometer and the hand-held goldmann applanation tonometer in pediatric aphakia

Yunping Li1, Luosheng Tang, Manyi Xiao

  • 1Department of Ophthalmology, 2nd Xiangya Hospital, Central South University, Hunan, China.

Journal of Glaucoma
|September 26, 2012
PubMed

Insights

The Icare tonometer is well-tolerated by pediatric aphakia patients, though it may overestimate intraocular pressure (IOP) compared to the Perkins tonometer, especially with increased central corneal thickness.

Area of Science:

  • Ophthalmology
  • Pediatric Ophthalmology
  • Medical Devices

Background:

  • Congenital cataract surgery can lead to aphakia, necessitating regular intraocular pressure (IOP) monitoring.
  • Accurate IOP measurement is crucial for managing glaucoma risk in pediatric aphakic eyes.
  • Traditional tonometry methods can be challenging in young children, often requiring sedation.

Purpose of the Study:

  • To compare intraocular pressure (IOP) measurements between the Icare tonometer and the hand-held Goldmann applanation tonometer (Perkins tonometer).
  • To evaluate the usability and tolerance of the Icare tonometer in children with aphakia post-congenital cataract surgery.

Main Methods:

  • A study involving 125 children with aphakia following congenital cataract surgery.
  • Patients were divided into younger (3-30 months) and older (31-72 months) age groups.
  • IOP, axial length, and central corneal thickness (CCT) were measured; younger children underwent measurements under sedation.

Main Results:

  • A strong positive correlation was found between IOP readings from the Icare and Perkins tonometers (r=0.943, P<0.001).
  • The 95% limits of agreement ranged from -1.6 to 5.6 mm Hg.
  • Icare IOP measurements increased disproportionately with increasing CCT (r=0.408, P<0.001), while axial length showed no correlation with IOP.

Conclusions:

  • The Icare tonometer is generally better tolerated in pediatric aphakic patients, particularly under unsedated conditions.
  • The Icare tonometer may overestimate IOP compared to the Perkins tonometer.
  • The discrepancy between Icare and Perkins IOP measurements increases with greater central corneal thickness.