Related Experiment Video
Updated: May 18, 2026

Near-Infrared Spectroscopy During Reactive Hyperemia for the Assessment of Lower Limb Vascular Function
Published on: March 22, 2024
Unresolved instrumentation problems following clinical trials using near infrared spectroscopy
Abstract:
Near infrared spectroscopy (NIRS) clinical trials conducted over a seven year period have identified instrument engineering problems related to fiber optic failure, electromagnetic interference, chromophore algorithms, and computational software. These problems have caused confusion amongst clinicians at the bedside, rejection of large volumes of data, repeated reanalysis of data, and a significant diversion of project resources away from clinical studies and into engineering solutions. This article summarizes previously published studies and presents new data which, together, emphasize the need for improvements in NIRS technology. Instrument designers need to be aware of the need for these improvements if NIRS is to serve clinicians better during research designed to rationally define clinical management protocols. © 1998 Society of Photo-Optical Instrumentation Engineers.