[Analysis of multi-micromirrors tilt error in space-modulated Fourier transform infrared spectrometry]
Ying Zheng1, Jing-Qiu Liang, Zhong-Zhu Liang
1State Key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China. opticszhy@163.com
Abstract:
The system of space-modulated Fourier transform infrared spectrometer without moving parts is stable and compact, moreover, it can realize the on-line monitoring and real-time detection. The position accuracy of multi-micromirrors as the key components has a great influence on spectrometric performance. In the present paper, the multi-mcromirrors tilt error caused by optical path difference changes was calculated and simulated. The analysis shows that, as the multi-micromirrors tilt angle increased, the interference pattern contrast decreased and the signal to noise ratio reduced. The results could be used in the tolerance distribution and the systemic alignment in the next step.
Related Concept Videos
IR Spectrometers
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to the...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
IR Spectroscopy: Molecular Vibration Overview
Stretching vibrations are vibrational motions that occur along the bond line, changing the bond length or distance between two bonded atoms. They are further distinguished as symmetric or asymmetric. In symmetric stretching, the...