Organ dose and effective dose estimation in paediatric chest radiographic examinations by using pin silicon
Toshio Kawasaki1, Takahiko Aoyama, Chiyo Yamauchi-Kawaura
1Department of Radiological Technology, Kanagawa Children's Medical Center, Mutsukawa 2-138-4, Minami-ku, Yokohama 232-8555, Japan. t-kawasaki@c3-net.ne.jp
Abstract:
Organ and effective doses during paediatric chest radiographic examination were investigated for various tube voltages between 60 and 110 kV at a constant milliampere-second value and focus-to-film distance by using an in-phantom dose measuring system and a Monte Carlo (MC) simulation software (PCXMC), where the former was composed of 32 photodiode dosemeters embedded in various tissue and organ sites within a 6-y-old child anthropomorphic phantom. Lung doses obtained ranged from 0.010 to 0.066 mGy and effective doses from 0.004 to 0.025 mSv, where these doses varied by a factor of 6 with the change in the tube voltage. Effective doses obtained using the MC simulation software agreed with those obtained using the dose measuring system within 23 %, revealing the usefulness of PCXMC software for evaluating effective doses. The present study would provide helpful dose data for the selection of technical parameters in paediatric chest radiography in Japan.