Related Experiment Video
Updated: May 18, 2026

The Application of 1% Methylene Blue Dye As a Single Technique in Breast Cancer Sentinel Node Biopsy
Published on: June 1, 2019
Sentinel lymph nodes fluorescence detection and imaging using Patent Blue V bound to human serum albumin
Franklin Tellier1, Jérôme Steibel, Renée Chabrier
1Laboratoire d'Imagerie et de Neurosciences Cognitives, UMR 7237, Université de Strasbourg/CNRS, Faculté de Médecine, 4 rue Kirschleger, 67085 Strasbourg, France ; EURORAD, 2 rue Ettore Bugatti 67201 Eckbolsheim, France.
Abstract:
Patent Blue V (PBV), a dye used clinically for sentinel lymph node detection, was mixed with human serum albumin (HSA). After binding to HSA, the fluorescence quantum yield increased from 5 × 10(-4) to 1.7 × 10(-2), which was enough to allow fluorescence detection and imaging of its distribution. A detection threshold, evaluated in scattering test objects, lower than 2.5 nmol × L(-1) was obtained, using a single-probe setup with a 5-mW incident light power. The detection sensitivity using a fluorescence imaging device was in the µmol × L(-1) range, with a noncooled CCD camera. Preclinical evaluation was performed on a rat model and permitted to observe inflamed nodes on all animals.