Reliability and validity of quantifying absolute muscle hardness using ultrasound elastography
Kentaro Chino1, Ryota Akagi, Michiko Dohi
1Department of Life Sciences (Sports Sciences), Graduate School of Arts and Sciences, the University of Tokyo, Tokyo, Japan. chinokentaro@idaten.c.u-tokyo.ac.jp
Plos One
|October 3, 2012
Summary
This study verified a new ultrasound elastography method for measuring absolute muscle hardness. The method proved reliable and valid for quantifying muscle stiffness in both materials and human subjects.
Area of Science:
- Biomechanics
- Medical Imaging
- Musculoskeletal Research
Background:
- Muscle hardness, a measure of transverse muscle stiffness, is crucial for understanding muscle function.
- Previous quantitative methods for muscle hardness using ultrasound elastography lacked complete reliability and validity verification.
Purpose of the Study:
- To rigorously verify the reliability and validity of a quantitative ultrasound elastography method for absolute human muscle hardness measurement.
- To assess the method's accuracy in both in vitro and in vivo settings.
Main Methods:
- Ultrasound elastography was used to quantify Young's moduli of tissue-mimicking materials (20-80 kPa) and the medial gastrocnemius muscle.
- A strain/Young's modulus ratio derived from two reference materials (7 and 30 kPa) was employed for calculations.
- Intra- and inter-investigator reliability was assessed using coefficient of variations and intraclass correlation coefficients.
Main Results:
- The method demonstrated high reliability with low coefficients of variation (≤6.9%) and high intraclass correlation coefficients (≥0.77).
- A strong correlation (0.996) was found between ultrasound elastography and mechanical methods for tissue-mimicking materials.
- In vivo measurements of medial gastrocnemius muscle Young's moduli aligned with previously reported magnetic resonance elastography values.
Conclusions:
- The reliability and validity of the quantitative ultrasound elastography method for absolute muscle hardness measurement are confirmed.
- This validated method offers a promising tool for non-invasive assessment of muscle mechanical properties.
More Related Videos
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
Endoscopic Ultrasound (EUS):
Ultrasonography
Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called a...
During an ultrasonography procedure, a handheld device called a...