Related Experiment Video
Updated: May 18, 2026

Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
Direct comparison of a Ca+ single-ion clock against a Sr lattice clock to verify the absolute frequency measurement
Kensuke Matsubara1, Hidekazu Hachisu, Ying Li
1National Institute of Information and Communications Technology, 4-2-1 Nukui-kitamachi, Koganei, Tokyo 184-8795 Japan. matubara@nict.go.jp
Abstract:
Optical frequency comparison of the (40)Ca(+) clock transition ν(Ca)((2)S(1/2-)(2D(5/2), 729 nm) against the (87)Sr optical lattice clock transition ν(Sr) ((1)S(0)-(3)P(0), 698 nm) has resulted in a frequency ratio ν(Ca) / ν(Sr) = 0.957 631 202 358 049 9(2 3). The rapid nature of optical comparison allowed the statistical uncertainty of frequency ratio ν(Ca) / ν(Sr) to reach 1 × 10(-15) in 1000s and yielded a value consistent with that calculated from separate absolute frequency measurements of ν(Ca) using the International Atomic Time (TAI) link. The total uncertainty of the frequency ratio using optical comparison (free from microwave link uncertainties) is smaller than that obtained using absolute frequency measurement, demonstrating the advantage of optical frequency evaluation. We note that the absolute frequency of (40)Ca(+) we measure deviates from other published values by more than three times our measurement uncertainty.