Risk of dislocation using large- vs. small-diameter femoral heads in total hip arthroplasty
Johannes F Plate1, Thorsten M Seyler, D Alex Stroh
1Department of Orthopaedic Surgery, Wake Forest School of Medicine,Medical Center Boulevard, Winston-Salem, NC 27157-1070, USA.
Background:
Dislocation remains a difficult problem in total hip arthroplasty. Large-diameter femoral heads may lower the incidence of dislocation by enhancing the jump distance and decreasing impingement, but their performance against small-diameter heads has not been assessed. This study compared the mid-term radiographic and functional outcomes of two matched cohorts of patients undergoing total hip arthroplasty who had a high pre-operative risk for dislocation and who received either small-diameter (26- or 28-millimeters) or large-diameter (≥36-millimeters) femoral heads.
Methods:
All patients who received large-diameter heads (≥36-millimeter) between 2002 and 2005, and who had pre-operative risk factors for dislocation, were identified in the institution's joint registry. Forty-one patients (52 hips) who received large-diameter heads were identified, and these patients were matched to 48 patients (52 hips) in the registry who received small-diameter femoral heads.
Results:
At mean final follow-up of 62 months (range, 49 to 101 months), both groups achieved excellent functional outcomes as measured by Harris Hip scores, with slightly better final scores in the large-diameter group (90 vs. 83 points). No patient showed any radiographic signs of loosening. No patient dislocated in the large-diameter femoral head group; the smaller-diameter group had a greater rate of dislocation (3.8%, 2 out of 52).
Conclusions:
Large-diameter femoral head articulations may reduce dislocation rates in patients who have a high pre-operative risk for dislocation while providing the same functional improvements and safety as small-diameter bearings.