The configuration of bi-level ventilator circuits may affect compensation for non-intentional leaks during
Annalisa Carlucci1, Annia Schreiber, Alessio Mattei
1Respiratory Intensive Care Unit, Fondazione S. Maugeri, IRCCS, Pavia, Italy. annalisa.carlucci@fsm.it
Purpose:
To assess the behaviour of a pressure-preset volume-guaranteed (V(TG)) mode of ventilation in the presence of non-intentional leaks in single-limb circuit (SLC) home ventilators.
Methods:
All SLC home ventilators commercially available in Italy can be used in a V(TG) mode with an intentional leak ("vented") or a true expiratory valve ("non-vented") configuration were selected. Using an experimental model consisting of a mannequin connected to an active lung simulator, for each level of leak (15, 25 and 37 l/min) three different conditions of respiratory mechanics (normal, restrictive and obstructive) were simulated using the ventilators in either a "vented" or "non-vented" configuration.
Results:
Three home ventilators were tested: Vivo50 (Breas), PB560 (Covidien) and Ventilogic LS (Weimann). In a "vented" circuit configuration all three ventilators kept constant or increased inspiratory pressure in all leak conditions to guarantee the V(TG). Conversely, in a "non-vented" circuit configuration, all tested ventilators showed a drop in inspiratory pressure in all leak conditions, resulting in a concomitant reduction in delivered tidal volume. The same behaviour was found in all conditions of respiratory mechanics. In the absence of leaks, all the ventilators, independently of circuit configuration, were able to maintain the set V(TG) in the presence of modifications of the respiratory mechanics.
Conclusions:
The ability of the V(TG) mode to compensate for non-intentional leaks depends strictly on whether a "vented" or "non-vented" circuit configuration is used. This difference must be taken into account as a possible risk when a V(TG) mode is used in the presence of non-intentional leaks.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: