Robustness of iCIEF methodology for the analysis of monoclonal antibodies: an interlaboratory study
Oscar Salas-Solano1, Babu Kennel, SungAe Suhr Park
1Analytical Biochemistry Department, Seattle Genetics Inc., Bothell, WA 98021, USA. osalas@seagen.com
An international team including 12 laboratories from 11 independent biopharmaceutical companies in the United States and Switzerland was formed to evaluate the precision and robustness of imaged capillary isoelectric focusing for the charge heterogeneity analysis of monoclonal antibodies. The different laboratories determined the apparent pI and the relative distribution of the charged isoforms for a representative monoclonal antibody sample using the same capillary isoelectric focusing assay. Statistical evaluation of the data was performed to determine within and between laboratory consistencies and outlying information. The apparent pI data generated for each charged variant peak showed very good precision between laboratories with RSD values of less than 0.8%. Similarly, the RSD for the therapeutic monoclonal antibody charged variants percent peak area values are less than 11% across different laboratories using different analyst, different lots of ampholytes and multiple instruments. These results validate the appropriate use of imaged capillary isoelectric focusing in the biopharmaceutical industry in support of process development and regulatory submissions of therapeutic antibodies.
An international team including 12 laboratories from 11 independent biopharmaceutical companies in the United States and Switzerland was formed to evaluate the precision and robustness of imaged capillary isoelectric focusing for the charge heterogeneity analysis of monoclonal antibodies. The different laboratories determined the apparent pI and the relative distribution of the charged isoforms for a representative monoclonal antibody sample using the same capillary isoelectric focusing assay. Statistical evaluation of the data was performed to determine within and between laboratory consistencies and outlying information. The apparent pI data generated for each charged variant peak showed very good precision between laboratories with RSD values of less than 0.8%. Similarly, the RSD for the therapeutic monoclonal antibody charged variants percent peak area values are less than 11% across different laboratories using different analyst, different lots of ampholytes and multiple instruments. These results validate the appropriate use of imaged capillary isoelectric focusing in the biopharmaceutical industry in support of process development and regulatory submissions of therapeutic antibodies.
More Related Videos
06:37Using Reference Reagents to Confirm Robustness of Cytokine Release Assays for the Prediction of Monoclonal Antibody Safety
Published on: September 15, 2023
10:50Purification and Analytics of a Monoclonal Antibody from Chinese Hamster Ovary Cells Using an Automated Microbioreactor System
Published on: May 1, 2019