Reproducibility of radionuclide left ventricular ejection fraction in patients awaiting cardiac transplantation

A M Keogh1, S Eberl, M G Yeates

  • 1Department of Cardiac Transplantation, St. Vincent's Hospital, Darlinghurst, Sydney, Australia.

Insights

Radionuclide left ventricular ejection fraction (LVEF) assessment shows lower reproducibility at very low levels in dilated cardiomyopathy patients. A change of 0.08 U is needed to confirm real changes in severe LV dysfunction.

Area of Science:

  • Cardiology
  • Nuclear Medicine
  • Biomedical Imaging

Background:

  • Left ventricular ejection fraction (LVEF) is crucial for assessing systolic function and managing dilated cardiomyopathy.
  • Current methods for LVEF assessment lack reported reproducibility data at extremely low levels (<0.30).

Purpose of the Study:

  • To evaluate the reproducibility of radionuclide-derived LVEF in patients with severe left ventricular dysfunction (LVEF < 0.30).

Main Methods:

  • Seventeen symptomatic dilated cardiomyopathy patients (NYHA Class III/IV) underwent radionuclide LVEF scans twice, 72 hours apart.
  • Two independent observers analyzed sequential scans to determine interobserver and interscan reproducibility.

Main Results:

  • Mean LVEF was 0.18 ± 0.06 (scan 1) and 0.17 ± 0.06 (scan 2).
  • Interobserver reproducibility (SD) was 0.03 U (R=0.76), interscan reproducibility (SD) was 0.03 U (R=0.62), and overall reproducibility (SD) was 0.04 U (R=0.50).
  • Interobserver variation (0.027 U) was lower than in normal volunteers (0.05 U).

Conclusions:

  • Radionuclide LVEF assessment demonstrates reduced reproducibility at levels below 0.30.
  • A change of ≥0.08 U is recommended to indicate a real change in LV function for LVEF < 0.30.
  • Clinicians must consider lower reproducibility when interpreting serial LVEF scans in severe LV dysfunction.